Total Human Apolipoprotein B
(Apo B) ELISA Assay
Intended Use:
To quantitate total human
Apolipoprotein B (Apo
B)
Principle of
Procedure:
Solid phase
capture sandwich ELISA assay using a
microwell format.
Shelf Life:
The expiration date for the package and each component is stated on the
label(s). Store all components at 2°- 8°
degrees C with the exception of the
standard, which should be stored at
-20oC .
Patient and Standard Dilutions:
Dilute each serum, plasma or
tissue culture fluid specimen to
be tested 1:1,000 by first diluting 1:100 in PBS and then making a
subsequent 1:10 dilution with the Apo B specimen diluent provided
to form a final dilution of 1:1,000.
Prepare serial two fold dilutions of the human Apo B standard: Neat (N), 1:2, 1:4,
1:8 etc. with the specimen diluent provided. Use the specimen diluent alone as
the blank control well.
Materials Supplied:
Anti-Human Apo B coated microwell strips 12 x 8 with
plastic frame
HRP conjugated affinity purified goat anti-Apo
B - 12mL
Apo B standard (pre-diluted 1:1000) - 1 ml
TMB/peroxide substrate color developer - 12mL
Apo B
specimen diluent - 60mL
Sulfuric acid termination reagent (0.5N) - 12mL
15
X Wash buffer
concentrate - 60mL
Limitations of the Procedure:
No single assay should be used as the only basis for
arriving at a diagnostic conclusion.
For research use only.
Dynamic Range:
2.11 mg/dL - 264mg/dL
Reproducibility:
C.V. 6% - 10% depending upon the region of the standard curve.
Assay Procedure:
*Allow each reagent to reach room temperature before use
1. Add
100uL of diluted specimen or standard to each microwell
2. Incubate
at room temperature for 45 minutes
3. Decant
and wash each microwell five times with wash buffer (dilute buffer 1:15 with
reagent grade water)
4. dd
100uL of HRP conjugated goat anti-Apo B to each well
5. Incubate
at room temperature for 45 minutes
6. Decant
and wash as in step 3
7. Add
100uL of TMB/peroxide substrate and incubate at room temperature for 15 minutes
8. Terminate
the reaction with 100uL of 0.18N sulfuric acid
9. Zero
the microwell reader at 450nm using the specimen diluent zero control well
10. Determine
the optical density (O.D.) of the remaining wells
11. Construct a standard curve using the O.D. values obtained for each of the standards
12. Interpolate the unknowns from the standard curve